Selecting the Right Evaporation Method for Small-Volume Sample Concentration

Nitrogen Blowdown vs. Centrifugal Vacuum Concentration 

When your laboratory workflow demands the concentration of several identical small-volume samples simultaneously, selecting the optimal evaporation technology can significantly impact processing efficiency, sample integrity, and overall operational costs. Two proven approaches dominate this application space: nitrogen blowdown evaporators (such as the Organomation MULTIVAP) and centrifugal vacuum concentrators (such as the CentriVap Vacuum Concentrator). Both technologies offer distinct advantages for parallel sample processing, and understanding their respective strengths will help you make an informed decision aligned with your specific analytical requirements.

 

Table of Contents

Understanding the Technologies

Processing Speed and Throughput

Sample Capacity and Batch Flexibility

Solvent Compatibility and Application Suitability

Cross-Contamination Risk and Sample Integrity

Temperature Control and Flexibility

Acquisition and Operating Costs

Maintenance Requirements and Longevity

Application-Specific Recommendations

Making the Decision: Key Selection Criteria

Conclusion

 

Understanding the Technologies

Nitrogen Blowdown Evaporation

Nitrogen blowdown systems work by directing a controlled stream of inert nitrogen gas across the surface of each sample, which reduces the vapor pressure above the liquid and accelerates solvent evaporation. The MULTIVAP series delivers nitrogen through individual stainless steel needles connected to a precision-engineered gas distribution manifold, with each sample receiving isolated gas flow. Heat is applied via either a heated aluminum dry block (providing temperatures from 30°C to 120°C) or a temperature-controlled water bath (30°C to 110°C), depending on the model configuration. The combination of gentle gas flow and controlled heating enables rapid, reproducible solvent removal while maintaining sample integrity.

 

Centrifugal Vacuum Concentration

Centrifugal vacuum concentrators like the CentriVap system combine three physical principles: centrifugal force, vacuum, and controlled heat. Samples are loaded into a rotor within a sealed chamber, where vacuum reduces the solvent boiling point while centrifugal force (up to 1725 rpm) prevents bumping and ensures that solvents boil from the surface downward. Microprocessor-controlled heating (ambient to 100°C) accelerates the evaporation process. Evaporated solvents are captured in a refrigerated cold trap, protecting both the sample and the vacuum pump from contamination.

 

Processing Speed and Throughput

Evaporation Rate Comparison

Direct head-to-head testing reveals notable differences in processing speed between these technologies. For methanol evaporation—one of the most common laboratory solvents—nitrogen blowdown demonstrates superior speed performance. In controlled comparisons at 35°C, the N-EVAP nitrogen evaporator completely evaporated 1.5 mL of methanol from 2 mL tubes in 30 minutes, while the SpeedVac centrifugal evaporator required 40 minutes for the same task. This represents a 33% faster evaporation rate for nitrogen blowdown in this specific application.

The speed advantage becomes even more pronounced when nitrogen evaporators operate at their recommended temperatures. At 63°C—the optimal bath temperature for methanol—nitrogen systems can achieve evaporation rates of 0.25 mL/min, over four times faster than rates observed in comparison studies. For other common solvents like acetonitrile, hexane, and dichloromethane, nitrogen blowdown consistently delivers faster drying times when operated at appropriate temperatures.

Centrifugal evaporators excel with certain high-boiling-point solvents and aqueous mixtures where their vacuum capability provides distinct advantages. The reduced boiling points achieved under vacuum (typically 2-6 mTorr for SpeedVac systems) enable efficient evaporation of water, DMSO, and DMF—solvents that would require excessively high temperatures or extended processing times with nitrogen blowdown alone.

 

Setup Time Considerations

One frequently overlooked factor affecting laboratory productivity is equipment setup time. Centrifugal evaporators impose a mandatory 45-minute warmup period before sample processing can begin, stemming from the refrigerated cold trap technology at the heart of these systems. This warmup penalty affects workflow scheduling and represents a fundamental design constraint. To minimize daily setup time, many laboratories maintain continuous electrical power to SpeedVac systems, keeping the refrigerated trap cold and ready for use.

Nitrogen blowdown systems offer immediate operational readiness. The MULTIVAP can begin heating while samples are being loaded, and a typical 15-minute warm-up to operating temperature allows evaporation to commence well before a centrifugal system completes its mandatory conditioning cycle. This operational flexibility translates directly to measurable productivity gains, particularly for laboratories processing multiple batches throughout the day.

 

Sample Capacity and Batch Flexibility

Nitrogen Blowdown Configuration Options

The MULTIVAP series offers multiple configurations optimized for different throughput requirements. The 48-position model features a 6×8 rectangular array with 1.0-inch center-to-center spacing, accommodating vials with outside diameters from 10-22 mm. For higher-capacity applications, the 80-position configuration provides an 8×10 array with 0.8-inch spacing for 10-17 mm diameter vials. Each system includes toggle switches for individual rows, enabling nitrogen conservation when processing partial batches.

Custom sample blocks are precision-machined to match specific vial dimensions, ensuring optimal heat transfer and uniform evaporation across all positions. This customization capability represents a significant advantage for laboratories working with specialized sample containers or requiring maximum heating efficiency.

 

Centrifugal Concentrator Capacity

The CentriVap Benchtop Concentrator features a 12-inch diameter chamber capable of holding up to 132 samples depending on rotor configuration. Various rotors accommodate different tube sizes, from microcentrifuge tubes to larger vessels, with specialized rotors available for 96-well microplates. The 6-Place Microtiter Plate Rotor can process up to 576 samples (six 96-well plates) in a single run, making it attractive for high-throughput screening applications.

Both systems excel at parallel processing, though they serve slightly different niches: nitrogen blowdown systems typically handle 30-80 individual tubes simultaneously, while centrifugal concentrators can process larger numbers of very small samples (microplate format) or moderate numbers of larger samples depending on rotor selection.

 

Solvent Compatibility and Application Suitability

Volatile and Potentially Hazardous Solvents

Nitrogen blowdown represents the preferred choice for volatile, potentially explosive solvents including hexane, heptane, diethyl ether, and tetrahydrofuran (THF). These solvents possess low boiling points and high vapor pressure, which can lead to rapid, uncontrolled evaporation and potential safety hazards in vacuum environments. The gentle concentration achieved through nitrogen flow at controlled temperatures minimizes explosion risk.

For enhanced safety when evaporating volatile solvents, the MULTIVAP offers an optional Z-Purge device that continuously purges the bath case with nitrogen or other inert gas, keeping stray solvent vapor away from heating elements. This added protection layer further reduces combustion risk during processing.

Chlorinated solvents such as chloroform and dichloromethane should generally be avoided in centrifugal evaporators due to potential reactivity with construction materials and safety concerns under vacuum conditions. Nitrogen blowdown systems handle these solvents routinely, with optional acid-resistant coatings available for particularly corrosive applications.

 

High-Boiling-Point Solvents

Centrifugal evaporation demonstrates superior performance with high-boiling-point solvents including DMSO, DMF, and water. The vacuum environment dramatically lowers the effective boiling point of these solvents, enabling efficient removal without exposing samples to potentially damaging high temperatures. Attempting to evaporate these solvents using nitrogen blowdown alone would require extended processing times or excessively high bath temperatures that might compromise sample integrity.

For laboratories regularly processing aqueous samples or working with metabolomics applications requiring DMSO or DMF removal, centrifugal concentrators provide a practical solution. The low-temperature operation (15-35°C typical for SpeedVac) protects heat-sensitive metabolites and volatile compounds.

 

 Cross-Contamination Risk and Sample Integrity 

Isolated vs. Shared Processing Environment

A critical distinction between these technologies lies in their sample processing environment. Nitrogen blowdown systems utilize isolated gas flow to each sample position, with individual needles delivering nitrogen directly onto each sample's surface. This design inherently minimizes cross-contamination risk between samples, as no shared vapor space exists.

Centrifugal evaporators process all samples within a shared, sealed chamber. While the centrifugal force helps contain samples within their tubes and reduces bumping, the potential for cross-contamination through the vapor phase exists, particularly with volatile analytes. Research has documented evaporation "crosstalk," where analytes carried out with volatilized solvents can migrate to adjacent wells, especially when concentration differences between samples are substantial.

Strategies to minimize centrifugal evaporator crosstalk include avoiding well overfilling (75% maximum capacity), starting with low vacuum and controlled heat, maintaining appropriate sample-to-surface distances, and pH adjustment for volatile analytes. Advanced plate adapters can provide additional protection. Despite these precautions, laboratories processing samples with dramatically different analyte concentrations or working with highly volatile compounds may find nitrogen blowdown's isolated-flow design advantageous.

 

Physical Stress on Samples

Nitrogen blowdown offers gentler sample processing that avoids the physical forces exerted during centrifugation. This characteristic proves particularly important for delicate samples that might be disturbed or damaged by spinning. Very small sample volumes (sub-milliliter) benefit from the more controllable evaporation that reduces loss risk compared to the vacuum and centrifugal forces within centrifugal systems.

Recent research comparing nitrogen blowdown, SpeedVac concentrator, and lyophilizer methods for cellular metabolome preservation found all three methods performed equally well in maintaining sample integrity. However, nitrogen blowdown distinguished itself through practical advantages including immediate operational readiness and superior temperature control flexibility.

 

Temperature Control and Flexibility

Nitrogen Blowdown Temperature Range

MULTIVAP systems offer extensive temperature control ranges that enable optimization for specific solvent characteristics. Dry block models provide digital temperature control from 30°C to 120°C with ±2°C accuracy, while water bath versions operate from 30°C to 110°C. This broad temperature capability allows laboratories to select the optimal evaporation conditions for diverse solvents and sample types.

The recommended approach sets bath temperature 2-3°C below the solvent's boiling point to achieve maximum concentration rates while preventing sample bumping or overheating. For methanol (boiling point 64-65°C), the recommended water bath temperature of 63°C delivers optimal performance. This targeted temperature control enables laboratories to balance evaporation speed with sample protection.

 

Centrifugal Concentrator Temperature Limitations

Standard CentriVap models provide microprocessor-controlled heating from ambient temperature to 100°C in 1-degree increments. However, SpeedVac units typically operate between 15-35°C for most applications, which significantly limits their evaporation efficiency compared to the higher temperatures achievable with nitrogen systems. This restricted operational range reflects the vacuum-based evaporation mechanism, where reduced pressure compensates for lower temperatures.

For applications requiring sub-ambient processing, refrigerated CentriVap models combine cooling systems (down to -4°C) with heating capability, enabling precise temperature control across a wide range. These specialized units serve applications demanding cold processing but represent a higher initial investment.

 

Acquisition and Operating Costs

Initial Equipment Investment

Cost comparisons reveal dramatic differences in equipment acquisition expenses. Research-based cost analysis shows nitrogen blowdown evaporators range from $2,000 to $15,000, while SpeedVac concentrators cost $10,000 to $30,000—often double to triple the nitrogen system investment. For a complete centrifugal evaporation setup including concentrator, cold trap, and vacuum pump, total costs can reach $40,000 or more.

The MULTIVAP's simpler mechanical design, with fewer complex components like vacuum pumps, refrigeration systems, and specialized rotors, contributes to its lower acquisition cost. This price advantage makes nitrogen blowdown particularly attractive for laboratories with limited capital budgets or those establishing new sample preparation capabilities.

 

Operating Expenses and Consumables

Nitrogen consumption represents the primary ongoing cost for blowdown systems. However, laboratories already maintaining nitrogen generators for mass spectrometry or other analytical instruments can split gas flow to the evaporator, effectively reducing consumable costs to near zero. Gas flow control at each sample position, combined with row-level shutoff switches, enables nitrogen conservation when processing partial batches.

For laboratories purchasing nitrogen cylinders, costs vary based on volume, purity requirements, and delivery frequency. Onsite nitrogen generation technology can dramatically reduce long-term gas expenses, with return on investment achievable within months to a few years depending on usage volume.

Centrifugal concentrators incur electricity costs for vacuum pump operation, refrigerated cold trap cooling, and centrifugal motor drive. Annual electric costs for compressed air and vacuum systems in comparable laboratory applications can exceed $100,000 for high-usage facilities, though individual concentrator operating costs are considerably lower. Vacuum pump oil requires regular monitoring and replacement when cloudy or contaminated, adding to maintenance costs.

 

Maintenance Requirements and Longevity

Nitrogen Blowdown System Maintenance

MULTIVAP systems require straightforward, minimal maintenance that laboratory personnel can easily perform. Stainless steel needles should be washed and dried after each use to prevent cross-contamination, using solvent wipes or laboratory washers (ultrasonic cleaner or autoclave). Water bath models require periodic cleaning to remove rust or scale buildup, with best practice recommending emptying the bath when not in use. For daily-use applications, air filter replacement occurs annually; with lower usage, filters can last five years or longer.

The simpler mechanical design of nitrogen evaporators, with fewer moving parts and no vacuum pumps or refrigeration systems, translates to reduced maintenance costs over the equipment's lifetime. Organomation backs MULTIVAP systems with one-year warranties and complimentary lifetime product support, with proper maintenance enabling instruments to last well over a decade.

 

Centrifugal Concentrator Maintenance

CentriVap systems demand more complex maintenance procedures involving multiple components. The refrigerated cold trap requires weekly draining of condensed solvents into appropriate waste containers, with ethanol replacement to prevent ice buildup in the stainless steel trap. Annual condenser cleaning using vacuum cleaners ensures proper airflow for peak performance. Acid cartridges (for acid-resistant models) require replacement when saturated.

Vacuum pump maintenance represents a significant ongoing requirement. Oil levels require frequent checking and topping when low, with complete oil changes necessary when cloudiness, particles, or discoloration appear. The useful life of vacuum pump oil extends when operators run pumps for extended periods after concentration runs, purging contaminants from hot oil. Some systems require defrosting and cleaning procedures that add downtime between runs.

Rotor assemblies have limited lifespans and represent expensive replacement components. High-speed operation subjects rotors to enormous forces, necessitating careful logging to track usage and ensure replacement before catastrophic failure. These maintenance demands require diligent training and operator discipline.  

 

Application-Specific Recommendations

Ideal Applications for Nitrogen Blowdown

Nitrogen blowdown evaporators excel in numerous analytical applications where they represent the gold standard for sample preparation. Environmental testing laboratories rely on nitrogen evaporation for EPA methods including 415.3, 515.2, 523.1, 525.3, 527, 537, 539, 548.1, 550, 1614, 1660, 1668, 1668A, 1668B, 1694, 1698, 8260E, 8280A, 8280B, 8290A, 8321B, and 625. These methods span applications from organic carbon determination and pesticide analysis to hormones, pharmaceuticals, and personal care products in drinking water and environmental samples.

The gentle, controlled evaporation provided by nitrogen systems proves essential for forensic chemistry applications where sample integrity and contamination prevention are paramount. Agricultural testing, food and beverage analysis, pharmaceutical quality assurance, and oil and grease determination all benefit from nitrogen blowdown's precision and reliability.

For laboratories processing volatile organic solvents (hexane, acetonitrile, dichloromethane, ethyl acetate, methanol) in moderate volumes (typically 1-15 mL per sample), nitrogen blowdown delivers optimal performance.

 

Ideal Applications for Centrifugal Concentration

Centrifugal vacuum concentrators represent the technology of choice for molecular biology, biotechnology, and proteomics laboratories processing DNA, RNA, and protein samples. The low-temperature vacuum environment protects biological activity and prevents thermal degradation of heat-sensitive biomolecules.

Metabolomics and lipidomics studies benefit from centrifugal evaporation's gentle processing and temperature control, with multiple samples processed uniformly in a controlled environment. Clinical diagnostic applications, particularly those involving 96-well microplate formats for hormone assays and drug screening, leverage the high-throughput capability of centrifugal systems.

Applications requiring removal of high-boiling-point solvents (DMSO, DMF, NMP) or significant water volumes find centrifugal concentration superior to nitrogen blowdown. The vacuum environment enables efficient processing without excessive heating or extremely long processing times.

For laboratories working predominantly with aqueous solutions, processing large numbers of very small samples (microplate format), or requiring overnight automated drying runs, centrifugal concentrators provide valuable capabilities.

 

Making the Decision: Key Selection Criteria

When evaluating these technologies for your laboratory, consider the following critical factors:

Solvent Profile: If your work emphasizes volatile organic solvents (hexane, acetonitrile, chlorinated solvents), nitrogen blowdown offers superior performance and safety. For high-boiling-point solvents or predominantly aqueous samples, centrifugal concentration provides advantages.

Sample Volume and Number: Both technologies handle multiple samples well, with nitrogen systems typically accommodating 30-80 individual tubes and centrifugal systems handling similar tube counts or up to 576 microplate wells. Assess your typical batch sizes and sample container formats.

Budget Constraints: Nitrogen blowdown requires 50-70% lower initial investment and reduced maintenance costs. Laboratories with tight capital budgets or establishing new capabilities may find this decisive.

Workflow Timing: If your laboratory requires immediate processing capability without warmup delays, nitrogen blowdown's instant readiness provides significant advantages. For overnight or planned runs where 45-minute warmup is acceptable, centrifugal systems work well.

Temperature Requirements: Applications demanding temperatures above 65°C benefit from nitrogen blowdown's extended range (up to 120°C). Low-temperature or sub-ambient requirements favor refrigerated centrifugal models.

Application Type: Environmental, forensic, agricultural, and petroleum testing traditionally employ nitrogen blowdown. Molecular biology, proteomics, and clinical diagnostics frequently use centrifugal concentration.

Cross-Contamination Sensitivity: For applications where sample-to-sample contamination represents a critical concern, nitrogen blowdown's isolated-flow design offers inherent protection.

 

Conclusion

Both nitrogen blowdown evaporation and centrifugal vacuum concentration represent proven, reliable technologies for parallel small-volume sample processing. The Organomation MULTIVAP nitrogen evaporator delivers immediate operational readiness, faster processing for volatile solvents, lower acquisition and maintenance costs, broader temperature control, and isolated gas flow that minimizes cross-contamination risk. The CentriVap Vacuum Concentrator excels with high-boiling-point solvents and aqueous samples, offers excellent performance for heat-sensitive biomolecules, provides high-capacity microplate processing, and enables low-temperature concentration under vacuum.

Rather than viewing these as competing technologies, consider them complementary tools serving different application niches. Many sophisticated laboratories maintain both systems, selecting the optimal technology based on specific sample characteristics, solvent properties, and workflow requirements. By carefully assessing your laboratory's predominant applications, sample types, budget parameters, and workflow demands, you can select—or combine—these technologies to optimize your sample preparation efficiency and analytical results.

Whether you choose the proven simplicity and speed of nitrogen blowdown or the vacuum-enhanced capability of centrifugal concentration, both the MULTIVAP and CentriVap represent excellent investments that will serve your laboratory's sample preparation needs for years to come.

 